|
CHAPTER 17. OPERATION OF EQUIPMENT
SECTION 9. HYGROTHERMOMETERS
17-76.
HYGROTHERMOMETER (HO-83, ASOS 1088, ETC.) -
GENERAL
Unlike previous hygrothermometers that employ a
lithium chloride dew cell to measure dew point, these hygrothermometers
use a "chilled mirror" technique. The sensor, in the
aspirator, contains a mirror and two temperature measuring beads. One
bead, in the intake of the aspirator, measures the ambient temperature.
Dew point temperature is determined by the mirror which is held at
precisely the temperature at which a thin film of condensation is
maintained on its surface. A thermal unit raises and lowers the
temperature of this mirror as necessary to maintain the thickness of the
condensation. The temperature of the mirror, measured by the other bead,
is the dew point temperature.
17-77.
HYGROTHERMOMETER (HO-83) -
OPERATIONS
Except for the possible cleaning of the mirror, if no
electronic technician is available at the station, the HO-83 contains no
operator serviceable parts. All instrument calibration and maintenance,
preventive and corrective, shall be done by an electronic technician.
However, the observer shall:
a. Be familiar with switches and
indicators present on the display.
b. Make a daily check of the
displays.
c. Compare hygrothermometer
against a standard as necessary but at least once a week.
17-78.
HYGROTHERMOMETER (HO-83) -
INSTRUMENT ERROR INDICATIONS
Malfunction of the internal circuits of the HO-83
will, in most cases, be indicated by:
a. The error light, which will
blink if no signal is being received or errors are suspected or
detected. The light will also blink for approximately 5 minutes
when power is restored after being shut off. It takes about
5 minutes for the computer to receive enough data to compute the
5-minute averages of temperature and dew point.
b. If no malfunction exists in
the display panel (pressing the display button), all displays
(temperature, dew point, maximum and minimum temperature) will read
-188.8
17-79.
HYGROTHERMOMETER (HO-83) -
QUALITY CONTROL
Equipment malfunctions will usually be indicated by
the error light or an erroneous display when the test button is pressed,
but equipment functioning properly can yield incorrect data if, for
example, the mirrors are dirty or the intake tube of the aspirator is
clogged. Although cleaning the mirrors and intake tube is usually done
by the electronic technician, the observer is responsible for ensuring
that the temperatures reported are valid by making periodic comparison
checks.
17-80.
HYGROTHERMOMETER (HO-83, ASOS
1088, ETC.) - COMPARISON CHECKS (NA LAWRS)
Normal comparison checks consist of comparing the
hygrothermometer's temperature and dew point with simultaneous values
obtained from a properly exposed comparison standard--a sling or shelter
mounted psychrometer. The comparison values may be observed at a site
convenient to the office as long as the values obtained are
representative of those at the hygrothermometer site. The observer shall
make comparisons when no ice fog is present, and:
a. Following preventive as well
as corrective maintenance.
b. At least once a week, as near
1200 LST on Monday as possible and more frequently when calibration is
suspect or previous comparisons suggest the possibility of unstable or
otherwise unsatisfactory performance. The frequency should be such as
to keep, as far as practical, the performance within the limits shown
below and in Figure 13-1, Chilled Mirror Hygrothermometers - Dew Point
Comparison Difference in ° F (e.g., HO-83,
1088, etc.), and in Figure 13-2, Chilled Mirror Hygrothermometers -
Dew Point Comparison Difference in ° C
(e.g., HO-83, 1088, etc.).
(1) For ambient temperatures,
the difference between the values obtained from the station standard
and the hygrothermometer shall not exceed 2°C.
This limit is applicable to comparisons made with the station
standard at either the hygrothermometer site or at the remote
location convenient to the office.
(2) Allowable dew point
differences are based on the temperature/dew point spread as
determined by the comparison standard (psychrometer). These are
shown in Figure 13-1, Chilled Mirror Hygrothermometers - Dew Point
Comparison Difference in ° F (e.g.,
HO-83, 1088, etc.), and in Figure 13-2, Chilled Mirror
Hygrothermometers - Dew Point Comparison Difference in °
C (e.g., HO-83, 1088, etc.), and, like temperature, are applicable
to comparisons at either comparison site.
(3) Valid persistent
temperature differences, if any, between the hygrothermometer site
and a remote site may be applied as a correction to the comparison.
(4) If limits are exceeded when
the comparison is made at the location remote from the
hygrothermometer site, the observer shall promptly make a comparison
at the site of the hygrothermometer. If limits are still exceeded,
the observer shall discontinue use of the system and notify the
electronic technician for corrective action.
(5) If limits are not exceeded
but the trend of comparisons for the past three weeks show a
relatively steady drift towards higher or lower temperature or
dew-point values, such as from a dirty mirror, the frequency of
comparisons should be increased.
(6) A graph of comparison
departures is required as an aid in evaluating system
performance and to determine the frequency of comparisons, NOAA Form
59-6 or other convenient graph paper may be used to plot plus or
minus departures above and below a horizontal zero reference line.
From the graph the need for corrective maintenance or for more
frequent comparisons should be readily apparent.
|