|
CHAPTER 17. OPERATION OF EQUIPMENT
SECTION 5. BAROMETER COMPARISONS
17-48. GENERAL
At locations with a mercury barometer, commissioned
ASOS, AWOS, or digital altimeter setting indicator (DASI) designated as
the station pressure standard, analog aneroid barometers and analog
altimeter setting indicators shall be standardized. Those analog
instruments shall be routinely compared to the station pressure standard
to ascertain continued reliability and to determine corrections to
readings of the analog aneroid instruments. Comparison data shall be
entered on MF1-13, using the guide for preparation on the back of the
form. Figure 17-10, Sample of MF1-13 Barometer Comparison (Front),
presents an illustration of the front side of Form MF1-13. Figure 17-11,
Sample of MF1-13, Barometer Comparison (Back), presents an illustration
of the back side of Form MF1-13. Station pressure standards shall be
compared with travelling standards that are directly traceable to
mercury standards at the National Institute of Standards and Technology
(NIST).
17-49. REQUIREMENTS FOR STANDARDIZING
INSTRUMENTS
A series of comparative readings, as specified in
paragraph17-50, Standardizing Procedures, shall be accomplished
whenever:
a. A station pressure standard is
replaced.
b. An aneroid barometer or
altimeter setting indicator is installed, reset, or relocated.
c. Erratic or excessive
corrections are observed (see paragraph 17-53, General).
17-50. STANDARDIZING PROCEDURES
When standardization of an aneroid instrument is
required (see Section 6, Performance of Aneroid Instruments, of
this chapter), the observer shall make comparisons and determine mean
corrections as follows:
a. Make 10 comparisons of the
aneroid instrument with the mercury barometer. If practicable, make
these readings at hour intervals but, in no case, should the interval
be less than 15 minutes.
b. During the time the 10
comparisons are being made, the station pressure and altimeter setting
for hour observations shall be based upon the readings of the mercury
barometer. The aneroid instrument, using the correction from the
preceding hour comparison, may be used to determine pressure values
for SPECI observations and local needs.
c. After the 10 comparisons have
been made, compute the mean of the column 10 entries. Check the
reliability of the instrument by verifying that the difference between
corrected readings does not exceed allowable differences (see
paragraph 17-65, Sling Psychrometer Ventilation). If determined
reliable, enter and use the mean correction in column 12 as the
"posted correction" to be applied to reading of the aneroid
instrument; i.e., until the mean correction is redetermined.
d. On each of the next
5 workdays, make two additional comparisons at a 6-hour interval
and determine the mean correction as in c above, using the last 10
entries in column 10; e.g., using comparisons 3 through 12 on the
first of the 5 days, 5 through 14 on the second day, etc.
Figure 17-10. Sample of MF1-13, Barometer Comparison
(Front)

Figure 17-11. Sample of MF1-13, Barometer Comparison
(Back)

e. After the comparisons above
have been completed, and none of the differences between corrected
readings exceed allowable differences specified in paragraph 17-54,
Aneroid Barometer or Altimeter Setting Indicator, the aneroid
instrument may be considered standardized and used for routine
pressure readings.
f. When an instrument has been
standardized in accordance with the above instructions, the adjustment
screw (aneroid barometer) or elevation scale (altimeter-setting
indicator) shall not be reset without authorization from the
appropriate headquarters or maintenance shop.
17-51. ROUTINE COMPARISONS
After an aneroid instrument has been standardized and
accepted as reliable in accordance with paragraph 17-50, Standardizing
Procedures, the observer shall make two comparisons at least six hours
apart on the same day of each week, and:
a. Following each comparison,
shall conduct a check for reliability according to procedures in
paragraph 17-49, Requirements for Standardizing Instruments.
b. At the time of the second
comparison, shall determine a mean of the last 10 correction values
(column 10).
c. Shall enter the mean
correction in column 12 and check for excessive drift (see paragraph
17-50).
d. Provided the instrument is
determined reliable, shall use the mean (column 12) as the new posted
correction to be applied to the aneroid instrument readings; i.e.,
until the mean correction is redetermined the following week.
17-52. DISPOSITION OF FORM MF1-13
a. Disposition at LAWRS (ONLY LAWRS).
MF1-13 forms shall be retained on station until reviewed by a
representative from NWS Regional Headquarters.
b. Disposition At Stations Other Than
LAWRS. (NA LAWRS). Unless advised otherwise by the regional
headquarters, the station shall retain the completed forms on station
until they are reviewed by a representative from the regional
headquarters. Forms more than 18 months old that have been
reviewed may be destroyed.
|