|
CHAPTER 14. PRESSURE SECTION 3. OBSERVING, DETERMINING AND REPORTING PROCEDURES AT LAWRS 14-48. APPLICABILITY (ONLY LAWRS) The provisions of this section are applicable only at LAWRS. The only pressure related measurement required at LAWRS is the altimeter setting. 14-50. DETERMINING ALTIMETER SETTING (ONLY LAWRS) The altimeter setting shall be determined by use of an altimeter setting indicator (ASI), a digital altimeter setting indicator (DASI), ASOS, AWOS, electronic pressure transducer, or a mercury barometer. These instruments shall be routinely compared and corrected as described in this section. The correction for each instrument used to report altimeter setting shall be determined. The latest correction, even if zero, shall be displayed on the instrument. The posted correction shall be added to the instrument's reading before reporting the altimeter setting. Do not change the value of the altimeter setting in the ASOS one-minute page to add a correction. This will turn off report processing, which can only be turned back on by an NWS technician. 14-51. ALTIMETER SETTING FROM ALTIMETER SETTING INDICATORS (ONLY LAWRS) Altimeter setting indicators may be digital or analog. If a digital instrument is used, the altimeter setting shall be determined by adding the posted correction to the reading. Some digital altimeter setting indicators including electronic pressure transducers do not require any corrections. If an analog instrument is used, the following procedures shall apply:
14-52. DETERMINING THE RELIABILITY OF ALTIMETER SETTING INDICATORS (ONLY LAWRS) The reliability of each altimeter setting indicator (ASI) shall be verified as follows:
14-53. DAILY COMPARISON WITH AN ADJACENT STATION (ONLY LAWRS) Locations without an approved pressure standard may compare their altimeter device against values obtained from an adjacent NWS office, FSS, or a LAWRS with an approved pressure standard provided:
The observer shall not use altimeter setting values from aneroid instruments when the difference exceeds .02 at precision approach locations or .05 at all other locations. (Note: LAWRS locations using an automated sensor, such as ASOS or AWOS, are not required to conduct a daily comparison of the sensor output with an adjacent station.) 14-54. ALTIMETER SETTING FROM A MERCURY BAROMETER (ONLY LAWRS) Altimeter settings may be obtained from a mercury barometer. The mercury barometer may be used to determine station pressure from which the altimeter setting may be determined. The procedures for using the mercury barometer to determine station pressure are given in paragraphs 17-34 through 17-38. The altimeter setting may be determined from station pressure by use of a pressure reduction computer, reduction constant, or table. 14-55. ALTIMETER SETTING FROM A DASI Altimeter settings may be obtained from a DASI through direct readings. Some DASI’s may have a correction posted from a travelling standard. 14-56. ELECTRONIC PRESSURE TRANSDUCERS OR OTHER DASI’S A DASI may be used to determine station pressure from which the altimeter setting may be determined. The altimeter setting may be determined from station pressure by use of a pressure reduction computer, reduction constant, table, or by direct readout. 14-57. - 14-60. RESERVED
|